MP21-16 - Flexible Ureteroscope reprocessing: are high-level disinfected reusable scopes as clean as we assume?

Saturday, September 22
2:00 PM - 4:00 PM
Location: Room 243

Introduction & Objective : Reusable flexible ureteroscopes (USC) are reprocessed for subsequent use. Inadequate reprocessing may lead to (cross-)contamination in subsequent patients. So far, little is known about the effectiveness of USC reprocessing, especially with regard to cumulative USC use. We hypothesize that the effectiveness of reprocessing declines with cumulative USC use as USC wear and tear can result in surface irregularities that might form a breeding ground for microorganisms. This study evaluates the prevalence of preoperative USC contamination with increased UCS use and the possibility of cross-contamination.


Methods : From December 2015 until December 2017, data on scope use and microbiological cultures were collected prospectively for 20 new USCs (Karl Storz FlexXc and X2, Olympus URF V2 and P6). High-level disinfection with peracetic acid was used for USC reprocessing. To assess pre- and post-operative contamination, two microbiological samples of the USC were taken before and after the procedure (1. stirring the distal USC shaft tip in 10 mL saline, 2. flushing the working channel with 10 mL of saline).All four samples were sent for culture on bacteria and yeasts. The possibility of cross-contamination was evaluated by comparing the post-operative culture with the subsequent preoperative culture of the same USC. A positive culture was defined by ≥30 CFU/mL of skin flora or ≥10 CFU/mL of uropathogens. A generalized estimating equation model was used to analyse if factors related to USC use (number of procedures, ureteroscopy time, laser energy and number of device passes through working channel) were associated with an increased probability of positive pre-operative cultures.


Results : In total 389 procedures were performed. The total number of procedures performed per USC ranged from 2-80. Preoperative cultures were positive in 47/389 (12.1%) procedures. Uropathogens were found in 9 of the 47 positive cultures. In one case, the preoperative culture contained the same bacteria type (S. Aureus) as in the prior postoperative culture of the same USC. None of the factors related to cumulative USC use were associated with a higher probability of positive pre-operative cultures.


Conclusions : Microbiological contamination of reprocessed USC was found in 12.1% of all procedures. Cumulative USC use was not associated with a higher probability of preoperative contamination. The findings of this study indicate that there are flaws in high-level disinfection reprocessing.

Jaap D. Legemate

MD
Department of Urology, AMC University Hospital, Amsterdam, the Netherlands.

Resident in urology, PhD foccussed on ureteroscopic stone and tumour treatment.

Presentation(s):

    Send Email for Jaap Legemate

    Guido Kamphuis

    Consultant Urologist
    AMC Hospital Amsterdam

    Presentation(s):

      Send Email for Guido Kamphuis

      Jan Erik Freund

      Department of Urology, AMC University Hospital, Amsterdam, the Netherlands.

      Presentation(s):

        Send Email for Jan Erik Freund

        Joyce Baard

        Urologist
        AMC University Hospital, Urology, Amsterdam, The Netherlands

        Presentation(s):

          Send Email for Joyce Baard

          Harry Oussoren

          Department of Microbiology and Infection prevention, AMC University Hospital, Amsterdam, the Netherlands.

          Presentation(s):

            Send Email for Harry Oussoren

            Ingrid Spijkerman

            Department of Microbiology and Infection prevention, AMC University Hospital, Amsterdam, the Netherlands.

            Presentation(s):

              Send Email for Ingrid Spijkerman

              Jean de la Rosette

              AMC University Hospital, Amsterdam, the Netherlands

              Presentation(s):

                Send Email for Jean de la Rosette


                Assets

                MP21-16 - Flexible Ureteroscope reprocessing: are high-level disinfected reusable scopes as clean as we assume?



                Attendees who have favorited this

                Please enter your access key

                The asset you are trying to access is locked. Please enter your access key to unlock.

                Send Email for MP21-16 - Flexible Ureteroscope reprocessing: are high-level disinfected reusable scopes as clean as we assume?